Guralp-systems CMG-5T Manual de usuario Pagina 11

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 25
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 10
Operator's Guide
The primary gain block of the 5T is normally set at × 1 (unity). If
higher sensitivity is required, the gain of the instrument can be altered
by inserting gain-setting resistors onto the power supply board of the
unit:
If G is the gain change you require, then the value of the resistors to
insert is given by the formula
R = 10000 / (G – 1)
Care must be taken when soldering these resistors to the circuit board,
as overheating the terminal could easily damage surrounding circuit
elements.
December 2009 11
Vista de pagina 10
1 2 ... 6 7 8 9 10 11 12 13 14 15 16 ... 24 25

Comentarios a estos manuales

Sin comentarios